Analytik Jena – multi EA 5100 for micro elemental analysis
The multi EA 5100 is a versatile and reliable analyzer designed for the precise determination of sulfur, nitrogen, chlorine, and carbon in solid, liquid, and gaseous samples. It combines exceptional sensitivity with robust performance, making it ideal for high-throughput, cost-effective analysis across all matrices. Whether in quality control, research, or contract laboratories, the multi EA 5100 delivers consistent results while ensuring ease of use and excellent service support.
Fast, Versatile, and Highly Sensitive Elemental Analysis
The multi EA 5100 enhances laboratory efficiency and cost-effectiveness, providing high-quality elemental analysis with rapid processing times. Liquids can be analyzed in just three minutes, while solid samples require only six minutes, ensuring a high daily throughput of up to 160 liquid samples and 90 solid samples. Its low consumable consumption and minimal maintenance requirements further reduce operating costs.
With fully automated C/N/S/X testing, switching between carbon, nitrogen, sulfur, and chlorine determination is seamless—no manual adjustments or instrument modifications required. The system’s unmatched detector diversity provides a broad range of applications, delivering accurate and reliable results for any elemental analysis task. Its modular design allows for customization and expansion, adapting to changing laboratory needs. The HiPerSens detection system ensures a wide linear measuring range with exceptionally low detection limits.
Advanced Automation for Cost-Effective Analysis
The multi EA 5100 features flexible automation to maximize sample throughput while minimizing costs per analysis. A comprehensive range of accessories further boosts efficiency, including:
- Automatic Boat Drive (ABD) – Features a cooled sample port for the automated introduction of liquids and solids. The integrated flame sensor ensures matrix-optimized combustion, preventing soot formation and enabling precise analysis, even for unknown samples.
- Multi Matrix Sampler (MMS) – Automates the introduction of up to 112 liquid samples or 35 solid samples, streamlining workflows for high-throughput laboratories.
- Autoinjector – Enables safe, semi-automated injection of liquid samples up to 100 μl, ideal for small batch analysis.
- LPG and Gas Analysis Modules – Three matrix-optimized gas sampling solutions are available:
- GSS Module – Handles pressurized and unpressurized gas samples.
- LPG 2.0 Module – Designed for direct dosing of pressurized liquefied gases.
- GSS/LPG Combi Module – Prevents cross-contamination by using separate sample branches for liquefied and pressurized gases.
A Comprehensive Solution for Industrial Applications
With its wide application range and intelligent features, the multi EA 5100 is the ideal solution for industrial control labs, contract laboratories, and regulatory authorities across the oil and gas, petrochemical, and chemical industries. Its capabilities support applications in process control, material testing, and waste management.
- Flame sensor technology for matrix-optimized sample digestion, even for complex matrices such as oils, VGO, polymers, and liquefied gases.
- An integrated method library, ensuring easy and efficient routine analysis.
By combining speed, automation, and precision, the multi EA 5100 sets a new standard for reliable and cost-effective elemental analysis.
The multi EA 5100 is a versatile, all-in-one solution for the analysis of sulfur, nitrogen, chlorine, and carbon in solids, liquids, and gases. Designed for maximum flexibility, it combines exceptional robustness and high sensitivity, ensuring precise and reliable results across all sample matrices.
With its cost-effective high-throughput capabilities, the multi EA 5100 is an ideal choice for quality control laboratories, research facilities, and contract analysis. It delivers efficient performance, minimal maintenance, and outstanding service, making it a trusted solution for laboratories seeking both reliability and productivity.
Versatile, Fast, and Sensitive – Enhance Your Measurement Precision
The multi EA 5100 offers a modular design, allowing laboratories to customize and expand their system based on individual requirements. The basic unit can be upgraded at any time, providing flexibility to integrate the necessary accessory modules for specific applications.
With double furnace technology, the analyzer seamlessly adapts to different sample matrices. Its easy-to-use tilting mechanism enables a quick switch between vertical and horizontal combustion in just a few minutes, ensuring optimal combustion conditions for every sample type.
Equipped with an advanced safety system, the multi EA 5100 guarantees secure operation and extends the lifespan of key components. Automatic monitoring and optimization of system parameters ensure trouble-free performance, while the auto-protection system shields the analyzer from damage caused by particles, aerosols, and liquids.
The flame sensor technology enables matrix-optimized combustion, preventing soot formation and ensuring accurate, reliable results in minimal time.
- Double furnace system – Combines vertical and horizontal combustion in one analyzer for maximum flexibility.
- Intuitive user interface – Self-explanatory menu navigation for effortless operation.
- Self-Check System – Automatic monitoring, adaptation, and regulation of critical system parameters.
- High-Performance Gasbox & Flow Management System – Ensures stable gas flow, optimal digestion, and prevents soot formation with an automatic leak check.
- Auto-Protection System – Enhances safe operation and prolongs component lifespan.
- Plug-and-Start Technology – Automatic detection of system configuration for seamless operation.
- Flame Sensor Technology – Enables fast, soot-free combustion, independent of sample quantity and matrix composition.
With its cutting-edge automation, high sensitivity, and robust design, the multi EA 5100 ensures maximum efficiency, precision, and reliability in elemental analysis.
Detector | Injectıon Volume | Sample Supply | Sample Pressure | Mın. measurement time | Parameters | Dıgestıon Mode | Gas supply | Sample quantıty | Lımıt of Detectıon | Measurement Range | |
---|---|---|---|---|---|---|---|---|---|---|---|
multi EA 5100 for micro-elemental analysis 450-300.011 |
S: UVFD S: coulometric titration N: CLD Cl: coulometric titration C: NDIR | liquids: 0 – 500 µL gases: 0 – 100 mL LPG: 0 – 50 µL |
horizontal and vertical | gas: up to 200 bar LPG: up to 35 bar |
3 min | TS, TN, TC, TOC, TIC, EC/OC, TCl, TICl, TOCl, AOX, EOX (depending on accessories) | High-temperature combustion, catalyst-free | Oxygen 4.5 and Argon 4.6 | solids: 0 – 110 mg | S (UVFD): 5 ppb S (coulometry): 600 ppb N: 10 ppb Cl: 50 ppb C: 100 ppb |
S (UVFD): 0 – 1 wt-% S (coulometry): 0 – 4 wt-% N: 0 – 1 wt-% Cl: 0 – 10 wt-% C: 0 – 100 wt-% |
Standards
ASTM D1552-IR, ASTM D3120, ASTM D3246, ASTM D3961, ASTM D4629, ASTM D4744, ASTM D4929-B, ASTM D5194, ASTM D5453, ASTM D5762, ASTM D5808, ASTM D6069, ASTM D6667, ASTM D7183, ASTM D7184, ASTM D7457, ASTM D7551, DIN 15486, DIN 38409-22, DIN 38414-17, DIN 38414-18, DIN 51408-2, DIN 51444, DIN EN 14077, DIN EN 1484, DIN EN 1485, DIN EN 16166, DIN EN 17178, DIN EN ISO 9562, EN ISO 16591, EN ISO 20846, EPA 1650, CEPA 9020, BEPA 9023, EPA 9076GB/T 11060.4GB/T 11061GB/T 15959GB/T 17674GB/T 18612GOST ISO 20846GOST R 52247IP 379IP 490, ISO 11480, ISO 8245JIS K 2541-6JIS K 2609NEN 6402NEN 6979NEN EN 14077OENORM 6614SH/T 0253SH/T 0657SH/T 0689UOP 779UOP 910UOP 936UOP 971UOP 981UOP 987-AUOP 988VDI 2465-2