We have distilled over 50 years of sample preparation expertise into a microwave digestion system that meets every requirement. Whether you’re working with battery materials, environmental samples, food and feed, chemicals, pharmaceuticals, or handling routine to highly demanding samples, Multiwave 5001 and Multiwave 3001 are the ideal solutions for any throughput or application need.
With a 300 °C temperature capability, even the most challenging samples are digested quickly and efficiently. Tool-free vessel and rotor handling—with twice the speed for vessel opening and closing—enhances productivity and reduces digestion costs. Featuring 600 pre-installed programs, full-text method search, ready-to-use recipes, and a smart door design, Multiwave 5001 and Multiwave 3001 guarantee a seamless and optimized workflow.
Microwave Digestion You Can Trust
These microwave digestion systems come with the longest track record in sample preparation and extensive experience in microwave digestion, ensuring reliable and precise results with every run. No matter what the future holds, you can count on optimal sample preparation for trace metal analysis.
Multiwave 5001 is designed for a wide range of applications, from acid digestion to extraction and beyond. It delivers maximum throughput and the highest temperatures, enhanced by SmartTemp 2.0, the latest in accurate and stable temperature measurement, ensuring unmatched repeatability and consistent quality.
Multiwave 3001 is your cost-effective and versatile choice, covering most applications with efficiency and reliability.
Empowered by smart technology
With Multiwave 5001 and Multiwave 3001, say goodbye to repetitive tasks and let the integrated guides, software, and optimized design handle them for you. Even with minimal user input, these instruments deliver the highest sample throughput of up to 64 vessels in a single run and provide crystal-clear results every time. Game-changing features like SmartVent technology, tool-free operation, streamlined workflows, and the most extensive portfolio of rotors and accessories make handling effortless. With Multiwave instruments, you can achieve peak digestion temperatures regardless of sample size, as overpressure is safely released.
One microwave digestion platform for every application
Multiwave 5001 and Multiwave 3001 enable sample preparation through acid digestion and leaching, supporting environmental analysis, food testing, and pharmaceutical research.
Multiwave 5001 goes even further, offering compatibility with multiple sample preparation techniques, including:
Solvent extraction
Evaporation
Sample drying
Synthesis
Each of these processes is faster, safer, cleaner, and more cost-efficient compared to traditional, conventionally heated methods.
The main acid digestion rotors of Multiwave 5001 and 3001:1
Rotor 24HVT
Rotor 41HVT
Rotor 20SVT
Rotor 8 NXF
Rotor 64MG5
Number of vessels
24
41
20
8
64
Volume
50 mL/80 mL
56 mL
50 mL
100 mL
5 mL
Material
PTFE-TFM
PTFE-TFM
PTFE-TFM
PTFE-TFM
Glass
HF resistance
Yes
Yes
Yes
Yes
No
Temperature control
Internal T in all positions / SmartTemp 2.02
SmartTemp 2.02
T in all positions
IR in 16 positions
Pressure control
SmartVent technology / SmartVent detection
p in all vessels
PTFE seal
Applications
Routine samples: Biological and environmental samples, EPA procedures, food, cosmetic and pharmaceutical samples
Harder to digest samples: including polymers, ceramics, petroleum products, and alloys
Most difficult samples
Micro samples up to 20 mg
Compatible with
Multiwave 3001, Multiwave 5001
Multiwave 5001
Multiwave 5001
Multiwave 5001
1Multiwave 5001 is backwards compatible, it can accommodate Rotor 16 and accessories of older Multiwave models.
²Multiwave 5001 only
Standards
ASTM
D4309
D5258
D5513
D5765
D6010
D7210
D7876
D7879
Chinese Pharmacopoeia (ChP)
2321 Determination of lead, cadmium, arsenic, mercury and copper
EN
13346
13656
13657
13804
13805
13806
14083
14084
14332
14385
14627
14902
15297
15411
15763
15764
15765
EN ISO
15587-1
15587-2
EPA
3015
3051
3052
3546
European Pharmacopoeia (Ph. Eur.)
5.20. Elemental impurities
GB
5009
T 24533
T 26125
T 36592
T 36593
T 38287
HJ
491
678
680
702
703
765
766
781
803
832
921
922
ICH
Q3D Guideline
JIS
K 0050
Japanese Pharmacopoeia (JP)
2.63 ICP-AES & ICPMS (2. Preatreatment of Sample)
US Pharmacopoeia (USP)
232 – Elemental impurities – Limits
233 – Elemental impurities – Procedures
Ek bilgi
Number of Vessels
20, 24, 41
Volume
50 mL/80 mL, 56 mL
Material
PTFE-TFM
HF resistance
Yes
Temperature control
Internal T in all positions / SmartTemp 2.0, SmartTemp 2.0
Pressure control
SmartVent technology / SmartVent detection
Applications
Routine Samples: Biological and environmental samples, EPA procedures, food, feed, aqueous samples
Harder to digest samples: including polymers, ceramics, petroleum products, and alloys