Analytik Jena – multi N/C 2300 Series
The Specialist – A combustion analyzer specifically designed for handling particle-rich samples and those available in very small volumes, ensuring accurate and reliable results even under challenging conditions.
Sophisticated Technologies for Particle-Rich Samples and Small Volumes – multi N/C 2300 Series
Challenging sample matrices and demanding applications require a robust and intelligent analytical system. The multi N/C 2300 series meets these demands with innovative design and advanced technology. Its septum-free direct injection system, using microliter syringes, is equipped with an efficient rinsing mechanism that effectively prevents blockages and sample carryover—even when handling particle-rich or emulsion-like samples.
At the heart of the system lies the wide-range focus radiation NDIR detector, which allows for an extensive measurement range from 0 to 30,000 mg/L TOC. This enables direct analysis without the need for sample dilution. The Easy Cal function ensures long-term calibration stability and precise standard recovery. Additionally, the VITA flow management system guarantees reliable performance by automatically compensating for gas flow fluctuations.
For automatic sample handling, the AS 60 autosampler streamlines the workflow with integrated acidification, purging, and magnetic stirring for effective homogenization of liquid samples. An optional 112-position rack for 2 mL HPLC vials further enhances sample capacity and throughput.
The system also supports TC and TOC analysis in solid samples. The dual furnace module provides catalytic combustion at up to 950 °C within the instrument, while the HT 1300 solids module offers catalyst-free combustion up to 1800 °C (with additives) in a ceramic combustion tube. Solid samples of up to 3 grams can be introduced horizontally in ceramic boats—no foil wrapping required. The solids module is available as part of the fully automated multi N/C 2300 duo or can be retrofitted with manual or automated sample introduction as needed.
For nitrogen analysis, the instrument can be optionally equipped with a ChD (electrochemical detector) or an advanced CLD (chemiluminescence detector), either integrated during initial configuration or added later as an upgrade.
With its flexibility, durability, and high-performance design, the multi N/C 2300 series is the ideal choice for laboratories working with complex sample matrices, low volumes, and strict analytical requirements.
TOC/TNb Analysis for Particle-Rich Samples and Specialized Applications
The multi N/C 2300 series is specifically designed for demanding applications involving particle-rich samples. Utilizing direct injection technology, these compact and robust analyzers eliminate sample carryover and prevent blockages—even when handling complex liquid matrices. Their modular design allows easy customization for solid sample analysis and the simultaneous measurement of total nitrogen (TNb), offering exceptional versatility.
With septum-free injection via microliter syringes, even the smallest sample volumes can be analyzed precisely—ideal for scientific and pharmaceutical research. For example, in the total nitrogen-based determination of protein content in vaccines, the system enables fast, accurate measurements without cross-contamination and within a broad concentration range.
-
Outstanding particle handling with no carryover or blockages in the dosing path
-
Direct analysis of undiluted samples, eliminating the need for time-consuming pre-dilution
-
New intuitive software with live results display and an optional 21 CFR Part 11-compliant pharma module
-
Reduced operator workload thanks to the VITA flow management system, ensuring long-term calibration stability
-
Automated workflows including acidification, purging, and sample homogenization
-
Five-year warranty on both the wide-range focus radiation NDIR detector and furnace components
With its reliable performance and flexible configuration, the multi N/C 2300 is the perfect solution for laboratories working with challenging samples and strict quality standards.
Product Variants at a Glance
The multi N/C 2300 series offers flexible and robust solutions for TOC and TNb analysis across a wide range of sample types. Each model is tailored for specific applications—from particle-rich liquids to complex solids and pharmaceutical-quality nitrogen determination.
multi N/C 2300
This versatile TOC/TNb analyzer is ideal for environmental applications involving samples with high particulate or oil content. Direct injection via microliter syringe ensures economical sample handling and makes it particularly suitable for research where sample volumes are limited.
-
Optional solids analysis
-
Non-catalytic high-temperature combustion with HT 1300
-
Acid digestion module for TIC in solids
-
Double furnace module for catalytic combustion
-
-
Optional nitrogen analysis
-
Simultaneous TOC/TNb measurement
-
Integrated ChD detector
-
CLD-300 module (retrofittable)
-
-
Applications
-
Wastewater monitoring in treatment facilities
-
Industrial wastewater and eluates
-
Trace analysis in small-volume research samples (e.g., soil extracts, ice cores)
-
multi N/C 2300 duo
Designed for high-load aqueous and solid samples, this variant integrates automated solids analysis and delivers efficient TOC/TNb analysis even in heterogeneous matrices.
-
Automated solids analysis
-
Non-catalytic combustion at up to 1800 °C (with additives)
-
Optional TIC analysis via acid digestion
-
-
Optional nitrogen analysis
-
Integrated ChD detector
-
Retrofittable CLD-300 detector
-
-
Applications
-
Wastewater and industrial effluent analysis
-
Environmental research (e.g., soil and water studies)
-
Complex solids: soils, sediments, contaminated sites, waste, construction debris
-
multi N/C 2300 N
This specialized model is optimized for total protein determination using catalytic high-temperature combustion and CLD detection. Ideal for pharmaceutical applications requiring high throughput and minimal sample volume.
-
High-level automation
-
Integrated AS 60 autosampler
-
112-position rack for 2 mL HPLC vials
-
Significant sample volume savings
-
-
Nitrogen analysis
-
Includes CLD-300 detector module as standard
-
-
Applications
-
Quality control of vaccines
-
TN determination in protein solutions for pharmaceutical production
-
Compliant with pharmacopoeia protein determination standards
-
| Detector | Injectıon Volume | Measurıng Time | Precısıon | Parameters | Dıgestıon Mode | Lımıt of Detectıon | Measurement Range | |
|---|---|---|---|---|---|---|---|---|
| multi N/C 2300 – Direct injection TOC analyzer 450-500.100-2 |
Focus Radiation NDIR | 10 – 500 µL | 3 – 5 min | 1 – 2 % | TC, TIC, TOC, DOC, NPOC, NPOCplus | Catalytic high temperature combustion up to 950°C | 50 µg/L | 0 – 30000 mg/L |
| multi N/C 2300 duo – Fully automated TOC analyzer for liquids and solids 450-500.104-2 |
Focus Radiation NDIR | 10 – 500 µL | 3 – 5 min | 1 – 2 % | TC, TIC, TOC, DOC, NPOC, NPOCplus | Catalytic high temperature combustion up to 950°C & Catalyst-free oxidation up to 1300°C | 50 µg/L | 0 – 30000 mg/L |
| multi N/C 2300 N – TN analyzer for aqueous protein solutions 450-500.103-2 |
Chemiluminescence Detector (CLD) | 10 – 500 µL | 3 – 5 min | 1 – 2 % | TN | High temperature combustion up to 950°C | 5 µg/L | 0 – 200 mg/L |
Standards
APHA SM 5310 A, B, C, D, ASTM D7573, ASTM D8083, ASTM G144, DIN EN 12260, DIN EN 13137, DIN EN 1484, DIN EN 15936, DIN EN ISO 20236, EP 2.5.33, 7B, EPA 415.1, EPA 415.3, EPA 9060A-1, ISO 10694, ISO 8245, JP 29, 7B, USP 1057, 7.2
“>